SG2165 SMU Platform

SG2165 SMU Platform

  • Integrate the functions of 5 instruments into one system.
  • Typical output sources with measurement accuracy of 0.02%, supporting DC/pulse output modes.
  • Pulse output mode with a minimum pulse width of 100 us and a rise time of 10 us.
  • High-precision source with pA-level resolution and support for multi-channel parallel current expansion.
  • Supports 2-wire and 4-wire resistance measurements.
  • Supports SCPI driver and commands.

Description

The SG2165 Precision Source-Measure Unit (SMU) platform is capable of four-quadrant operation,accurately providing voltage or current outputs while simultaneously measuring voltage, current, and resistance, among other functions. It integrates the capabilities of a six-and-a-half digit digital multimeter (DMM), a five-and-a-half digit precision voltage source, current source, electronic load, and pulse generator.It is feature-rich, compact in size, and comes with standard testing interfaces, making it highly suitable for integration into test fixtures. The SG2165 Source-Measure Unit platform is primarily used for IVR testing in industries such as semiconductors, sensors, and modules. It is customized for production line testing and provides efficient and cost-effective testing and measurement solutions for automated ICT and FCT in production lines.

sg2165 smu platform

Product features

  • Combines the functions of 5 instruments (digital multimeter, voltage source, current source, electronic load, and pulse generator) into one unit.
  • Typical output sources with measurement accuracy of 0.02%, supporting DC/pulse output modes.
  • Pulse output mode with a minimum pulse width of 100 us and a rise time of 10 us.
  • High-precision source with pA-level resolution and support for multi-channel parallel current expansion.
  • Supports 2-wire and 4-wire resistance measurements.
  • Supports SCPI driver and commands.

 

I-V Testing Performance

sg2165 smu platform

 

Application Scope

Primarily used for IVR testing in industries such as semiconductors, sensors, and modules.

It is customized for production line testing and provides efficient and cost-effective testing and measurement solutions for automated ICT and FCT in production lines.

sg2165 smu platform

 

Specifications

Model SG2165
Specifications 20 W
Channels 1
Quadrants 4
Voltage Accuracy Range Setting Resolution Setting Accuracy Measurement Resolution Measurement Accuracy
200 mV 2 uV 0.02%+375 uV 15 nV 0.02%+375 uV
2 V 20uV 0.02%+600 uV 150 nV 0.02%+600 uV
20 V 200uV 0.02%+5 mV 1.5 uV 0.02%+5 mV
Current Accuracy Range Setting Resolution Setting Accuracy Measurement Resolution Measurement Accuracy
100 nA 1 pA 0.06%+50 pA 10 fA 0.06%+50 pA
1 uA 10 pA 0.03%+700 pA 100 fA 0.03%+700 pA
10 uA 10 0pA 0.03%+5 nA 1 pA 0.03%+5 nA
100 uA 1 nA 0.03%+60 nA 10 pA 0.03%+60 nA
1 mA 10 nA 0.03%+300 nA 100 pA 0.03%+300 nA
10 mA 100 nA 0.03%+6 uA 1 nA 0.03%+6 uA
100 mA 1 uA 0.03%+30 uA 10 nA 0.03%+30 uA
1 A 10 uA 0.05%+1.8 mA 100 nA 0.05%+1.8 mA
3 A (≤6 V ) 100 uA 0.4%+8 mA 1 uA 0.4%+8 mA
Output Voltage Source Range Typical Output Settling Time Test Conditions
200 mV <20 us Voltage Change Time at No Load, 10%-90%
2 V <50 us
20 V <200 us
Current Source 100 nA <10 ms Current Change Time at Output Short Circuit,10%-90%
1 uA <5 ms
10 uA <5 ms
100 uA <1 ms
1 mA <300 us
10 mA <20 us
100 mA <20 us
1 A <20 us
Voltage Ripple and Noise (10 Hz-20 MHz) 3 mVrms, 20 V range